Public test session report

Extrudr BioFusion Blue Fire PLA Blue

MaterialID MAT0115

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-13 19:31:28

Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.

Measurement provenance

Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.

ModuleMeasured date
TensileNot recorded
ImpactNot recorded
Izod2026-10-05
Charpy2026-10-05
StiffnessNot recorded
Rigid test status
Complete
Rigid result modules
6/6
Rigid specimen/result records
37
Public details
Approved

Rigid specimen and result quality

ModuleOrientationAverageStd. dev.CVSpecimensConfidenceValidation
TensileUpright1,252 MPa0,287 MPa0,229%77/10OK
TensileFlat33,115 MPa9,785 MPa0,295%1010/10OK
ImpactUpright7,481 kJ/m²0,65 kJ/m²0,087%1010/10OK
ImpactFlat17,073 kJ/m²4,03 kJ/m²0,236%99/10OK
StiffnessThree-point bend2527,565 MPan/an/a1CompleteOK

Recorded native inputs

ModuleInput setRecorded values
TensileUpright force inputs (N)9, 5, 10, 9, 7, 7, 10, , ,
TensileFlat force inputs (N)209, 200, 185, 186, 248, 219, 356, 236, 213, 101
ImpactUpright needle inputs (%)9,00, 9,00, 9,00, 8,00, 8,00, 10,00, 10,00, 10,00, 10,00, 9,00
ImpactFlat needle inputs (%)22,00, 25,00, 17,00, 31,00, 14,50, 19,00, 17,00, 21,00, 21,00,
StiffnessRevolutions / degrees2 / 10

Approved test notes

Method and equipment context

Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.

Testing methodology · Engineering methodology whitepaper · Download PDF

Izod and Charpy Impact Strength

Charpy Impact Strength · 2026-10-05
Instrument readings in kJ/m²
Mean 42.195 kJ/m²; Std. Dev 6.38 kJ/m²; CV 15.119%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.

Izod Impact Strength · 2026-10-05
Instrument readings in kJ/m²
Mean 27.444 kJ/m²; Std. Dev 6.994 kJ/m²; CV 25.486%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.