Public test session report

Fillamentum FX256 PA12 Clear

MaterialID MAT0022

3DPIceland Labs
Generated with 3DPIceland Engineering Platform v70.0.4 DIRECT-IMPACT-METHODOLOGY - Izod and Charpy Methodology
2026-10-13 19:31:28

Traceability boundary: canonical per-module measured dates are shown below. Session identifier, time of day, operator, printer/slicer profile and environmental conditions are Not recorded and are never inferred.

Measurement provenance

Canonical SQLite measured dates use ISO format; missing dates remain Not recorded.

ModuleMeasured date
TensileNot recorded
ImpactNot recorded
Izod2026-10-07
Charpy2026-10-07
StiffnessNot recorded
Rigid test status
Complete
Rigid result modules
6/6
Rigid specimen/result records
39
Public details
Approved

Rigid specimen and result quality

ModuleOrientationAverageStd. dev.CVSpecimensConfidenceValidation
TensileUpright15,704 MPa4,843 MPa0,308%1010/10OK
TensileFlat28,901 MPa2,545 MPa0,088%1010/10OK
ImpactUpright8,019 kJ/m²4,526 kJ/m²0,564%1010/10OK
ImpactFlat37,254 kJ/m²11,089 kJ/m²0,298%88/10OK
StiffnessThree-point bend768,801 MPan/an/a1CompleteOK

Recorded native inputs

ModuleInput setRecorded values
TensileUpright force inputs (N)117, 116, 131, 57, 144, 123, 85, 119, 60, 69
TensileFlat force inputs (N)182, 193, 172, 184, 195, 197, 196, 208, 201, 151
ImpactUpright needle inputs (%)9,00, 9,50, 8,00, 25,00, 6,00, 6,50, 9,50, 8,00, 10,00, 7,00
ImpactFlat needle inputs (%)35,00, 62,00, 23,00, 67,00, 49,00, 46,00, 70,00, 37,00, ,
StiffnessRevolutions / degrees6 / 240

Approved test notes

Method and equipment context

Purpose-built tensile, pendulum impact and three-point bend apparatus are used with the governed Settings Manager constants consumed by ResultsService. Results are comparative and do not claim accredited-laboratory equipment.

Testing methodology · Engineering methodology whitepaper · Download PDF

Izod and Charpy Impact Strength

Charpy Impact Strength · 2026-10-07
Instrument readings in kJ/m²
Mean 9.334 kJ/m²; Std. Dev 1.074 kJ/m²; CV 11.507%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.

Izod Impact Strength · 2026-10-07
Instrument readings in kJ/m²
Mean 7.858 kJ/m²; Std. Dev 0.698 kJ/m²; CV 8.878%; Samples 10; Confidence 10
No break 0; invalid 0; excluded 0.
Comparative measurements; standard conformity unverified.